Skip to main content

Table 6 Sensitivity and Specifity for position of the three diaphyseal screws (overlapping of 2nd cortical bone) in relation to CT (gold standard)

From: Comparison of a ceiling-mounted 3D flat panel detector vs. conventional intraoperative 2D fluoroscopy in plate osteosynthesis of distal radius fractures with volar locking plate systems

 

True positive

False negative

True negative

False positive

Sensitivity (%)

Specifity (%)

2D Fluoroscopy

16

4

24

5

80

83

“Movie”

19

0

30

2

100

94

CBCT

18

1

34

0

95

100