Skip to main content
Fig. 5 | BMC Musculoskeletal Disorders

Fig. 5

From: Test-retest reliability and concurrent validity assessment of a novel high-frequency sensor device for anterior tibial translation measurement in loaded and unloaded condition: an exploratory cross-sectional study

Fig. 5

Bland & Altmann plot shows the mean difference of the Lachmeter® and the new device during the Lachman test setting. The systematic bias of - 0.18 mm (95% CI -0.71, 0.34) marked as purple line, whereas the blue and yellow lines show the limits of agreement (LOA) and 95% CI of the upper 3.01 mm and lower - 3.38 mm limits (± 1.96 SD), respectively

Back to article page